| Gama2 /frequency | 24 hours3 | 90 days | 1 ano | 2 years | temperature coefficient/℃ 4 |
| Tensão DC | TCAL±1℃ | TCAL±5℃ | TCAL±5℃ | TCAL±5℃ |
| 100 mV | 0.0030+0.0030 | 0.0040+0.0035 | 0.0050+0.0035 | 0.0065+0.0035 | 0.0005+0.0005 |
| 1 V | 0.0020+0.0006 | 0.0030+0.0007 | 0.0040+0.0007 | 0.0055+0.0007 | 0.0005+0.0001 |
| 10 V | 0.0015+0.0004 | 0.0020+0.0005 | 0.0035+0.0005 | 0.0050+0.0005 | 0.0005+0.0001 |
| 100 V | 0.0020+0.0006 | 0.0035+0.0006 | 0.0045+0.0006 | 0.0060+0.0006 | 0.0005+0.0001 |
| 1000 V | 0.0020+0.0006 | 0.0035+0.0010 | 0.0045+0.0010 | 0.0060+0.0010 | 0.0005+0.0001 |
| Tensão AC de verdadeiro valor eficaz (True RMS)2,5,6 |
| Gamas de 100 mV, 1 V, 10 V, 100 V e 750 V |
| 5 Hz - 10 Hz | 0.35+0.02 | 0.35+0.03 | 0.35+0.03 | 0.35+0.03 | 0.035+0.003 |
| 10 Hz-20 KHz | 0.04+0.02 | 0.05+0.03 | 0.06+0.03 | 0.07+0.03 | 0.005+0.003 |
| 20 KHz-50 KHz | 0.10+0.04 | 0.11+0.05 | 0.12+0.05 | 0.13+0.05 | 0.011+0.005 |
| 50 KHz-100 KHz | 0.55+0.08 | 0.60+0.08 | 0.60+0.08 | 0.60+0.08 | 0.060+0.008 |
| 100 KHz-300 KHz | 4.00+0.50 | 4.00+0.50 | 4.00+0.50 | 4.00+0.50 | 0.200+0.020 |
| Resistência7 | Testing current |
| 100 Ω | 1 mA | 0.0030+0.0030 | 0.008+0.004 | 0.010+0.004 | 0.012+0.004 | 0.0006+0.0005 |
| 1 kΩ | 1 mA | 0.0020+0.0005 | 0.008+0.001 | 0.010?0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 10 kΩ | 100 μA | 0.0020+0.0005 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 100 kΩ | 10 μA | 0.0020+0.0005 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 1 MΩ | 5 μA | 0.002+0.001 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0010+0.0002 |
| 10 MΩ | 500 nA | 0.015+0.001 | 0.020+0.001 | 0.040+0.001 | 0.060+0.001 | 0.0030+0.0004 |
| 100 MΩ | 500 nA‖10 MΩ | 0.300+0.010 | 0.800+0.010 | 0.800+0.010 | 0.800+0.010 | 0.1500+0.0002 |
| DCcurrent | Queda de tensão da resistência interna |
| 100 μA | <0.11V | 0.010+0.020 | 0.040+0.025 | 0.050+0.025 | 0.060+0.025 | 0.0020+0.0030 |
| 1 mA | <0.11V | 0.007+0.006 | 0.030+0.006 | 0.050+0.006 | 0.060?0.006 | 0.0020+0.0005 |
| 10 mA | <0.05V | 0.007+0.020 | 0.030+0.020 | 0.050+0.020 | 0.060+0.020 | 0.0020+0.0020 |
| 100 mA | <0.5V | 0.010+0.004 | 0.030+0.005 | 0.050+0.005 | 0.060+0.005 | 0.0020+0.0005 |
| 1 A | <0.7V | 0.050+0.006 | 0.080+0.010 | 0.100+0.010 | 0.120+0.010 | 0.0050+0.0010 |
| 3 A | <2.0V | 0.180+0.020 | 0.200+0.020 | 0.200+0.020 | 0.230+0.020 | 0.0050+0.0020 |
| 10 A8 | <0.5V | 0.050+0.010 | 0.120+0.010 | 0.120+0.010 | 0.150+0.010 | 0.0050+0.0010 |
| Capacidade15 |
| 1.0000 nF | 0.50+0.50 | 0.50+0.50 | 0.50+0.50 | 0.50+0.50 | 0.05+0.05 |
| 10.000 nF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 100.00 nF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 1.0000 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 10.000 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 100.00 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| Gama2/frequency | 24hours3 | 90days | 1year | 2years | temperature coefficient/℃4 |
| TCAL ± 1℃ | TCAL ± 5℃ | TCAL ± 5℃ | TCAL ± 5℃ |
| Corrente AC de verdadeiro valor eficaz (True RMS)2,6,9 | Queda de tensão da resistência interna |
| Gamas de 100 μA, 1 mA, 10 mA e 100 mA | <0.011,<0.11, <0.05,<0.5V |
| 3 Hz - 5 kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.015+0.006 |
| 5-10kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.030+0.006 |
| (typicalvalue) |
| 1Arange | <0.7V |
| 3Hz-5kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz (valor típico) | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.030+0.006 |
| 3 A Range | <2.0V |
| 3 Hz - 5 kHz | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.030+0.006 |
| (Typicalvalue) |
| 10 A Range8 | <0.5V |
| 3 Hz - 5 kHz | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz (valor típico) | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.030+0.006 |
| Connectivity |
| 1 kΩ | 0.002+0.030 | 0.008+0.030 | 0.010+0.030 | 0.012+0.030 | 0.0010+0.0020 |
| DiodeTesting10 |
| 5 V | 0.002+0.030 | 0.008+0.030 | 0.010+0.030 | 0.012+0.030 | 0.0010+0.0020 |
| DCrate(Typicalvalue) |
| (Precisão de entrada normalizada) + (precisão de referência normalizada) |
| Temperatura11 |
| PT100 (DIN/IEC751) | Theprobe accuracy+0.05℃ |
| 5 kΩ thermistor | Theprobe accuracy+0.1℃ |
| Frequência: precisão ± (% de leitura)12.13 |
| Gamas de 100 mV, 1 V, 10 V, 100 V e 750 V 14 |
| 3 Hz-10 Hz | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
| 10 Hz-100 Hz | 0.03 | 0.03 | 0.03 | 0.03 | 0.035 |
| 100 Hz-1 KHz | 0 | 0.008 | 0.01 | 0.01 | 0.015 |
| 1 KHz-300 KHz | 0.002 | 0.006 | 0.01 | 0.01 | 0.015 |
| Onda quadrada15 | 0.001 | 0.006 | 0.01 | 0.01 | 0.015 |
| Erro adicional do tempo de porta ± (% de leitura)13 |
| Frequência | 1second | 0.1second | 0.01second |
| 3 Hz-40 Hz | 0 | 0.2 | 0.2 |
| 40 Hz-100 Hz | 0 | 0.06 | 0.2 |
| 100 Hz-1 KHz | 0 | 0.02 | 0.2 |
| 1 KHz-300 KHz | 0 | 0.004 | 0.03 |
| Onda quadrada15 | 0 | 0 | 0 |
1. Em DC: as especificações são válidas após 60 minutos de aquecimento, com o período de integração a 10 ou 100 NPLC e o zero automático ativado. Em AC: as especificações são válidas após 60 minutos de aquecimento, com o filtro AC lento e onda sinusoidal.
2. Todas as gamas têm 20 % de sobregama, exceto 1000 V DC, 750 V AC, 10 A DC, 3 A AC, 10 A AC e o teste de díodos.
3. Relativamente ao padrão de calibração
4. Fora da gama TCAL ± 5 ℃, cada grau (℃) acrescenta um coeficiente.
5. As especificações são válidas com entrada sinusoidal > 5 % da gama e > 1 mV RMS. A gama de 750 V AC está limitada a 8 × 10⁷ V·Hz.
Para entradas entre 1 % e 5 % da gama e frequência < 50 kHz, acrescente um erro de 0,1 % da gama. Entre 50 kHz e 100 kHz, acrescente 0,13 % da gama.
6. Resposta a baixa frequência: há três filtros, de 3 Hz, 20 Hz e 200 Hz. Acima da frequência do filtro escolhido não há erros adicionais.
7. As especificações são válidas para medições de resistência a 4 ou 2 fios (com o offset a zero). Sem a função de zero matemático, a medição a 2 fios terá um erro adicional de 0,2 Ω.
8. A gama de 10 A só está disponível no conector frontal. Cada ampere acrescenta 2 mA de corrente base, ou corrente de entrada > 5 A RMS.
9. As especificações são válidas com entrada sinusoidal > 1,5 % da gama e > 10 μA AC. 10. As especificações referem-se à tensão medida na entrada; a corrente de teste típica é de 1 mA e as suas variações alteram a queda de tensão na junção do díodo.
11. A sonda escolhida limita a gama e o erro de medição reais. A precisão da sonda inclui todos os erros de medição e de conversão de temperatura ITS-90. O R0 da PT100 pode ser ajustado para 100 Ω ± 5 Ω para eliminar o erro próprio da sonda.
12. Salvo indicação em contrário, as especificações são válidas após 60 minutos de aquecimento e com entrada sinusoidal, para um tempo de porta de 1 s (7 dígitos).
13. Válido para entradas sinusoidais e quadradas ≥ 100 mV. Para entradas de 10 mV a 100 mV, multiplique a % de erro de leitura por 10.
14. A gama é de 10 % a 120 % abaixo de 750 V AC.
15. A entrada de onda quadrada é especificada entre 10 Hz e 300 kHz.